Chapter 6: troubleshooting – Labconco WaterPr PS Systems 90007 Series Manuel d'utilisation

Page 34

Advertising
background image

Chapter 6: Troubleshooting

30

Original Instructions

PROBLEM

CAUSES

CORRECTIVE

ACTION

Recirculation
pump operates but
no water is
dispensed (cont.)

Return line check valve is

defective

Replace check valve assembly.

Recirculation pump is defective

Replace recirculation pump.

Reduced flow at
dispense valve/gun

Optional final filter is blocked

Replace final filter.

Restricted tube in the feed line or

polishing system

Inspect polishing loop tubing for

any restrictions.

Water quality
display acting
erratically

New cartridges installed in

system

Purge all of the air out of the

polishing system.

Air trapped in resistivity cell

Purge all of the air out of the

polishing system.

Deionization

cartridges

are

exhausted

Replace

the

deionization

cartridges.

Polishing loop resistivity cell not

connected to wiring harness

Reconnect cell to wiring harness.

Polishing loop resistivity cell is

defective

Replace

cell.

Polished water
will not rinse up to
desired quality

Cartridges are past expiration

date

The shelf life of unopened

cartridges is two years from the
date of manufacture. Replace the
expired cartridges with new ones.
Check the manufactured date.

Cartridges installed in wrong

sequence

Install the cartridges as described

in the “Initial Operation and
Cartridge Replacement” section.

Water is bypassing the cartridges Ensure that the O-rings are at the

top of the deionization cartridges
and the posts at the top of the
bowls are inserted correctly.