Appendix c: waterpro ps system specifications – Labconco WaterPr PS Systems 90007 Series Manuel d'utilisation

Page 44

Advertising
background image

Appendix C: WaterPro PS System Specifications

40

Original Instructions

Water Dispensing Systems:

GUN

DISPENSING

MODELS

Dispense from gun by depressing trigger or from
dispense valve by pressing dispense key. Release
trigger or key to stop flow. Timed dispense from
dispense valve only. Optional hollow fiber filter can
be installed on both gun and dispense valves by
removing threaded nozzle and replacing with
hollow fiber filter.

NON-GUN DISPENSING MODELS
Dispense by pressing dispense key. Release key to
stop flow.

Water Quality Produced:

Meets or exceeds the following:

American Society for Testing and Materials
Type I Water

National Committee for Clinical Laboratory
Standards Type I Water

*Actual flow rates for ultrafiltered models could vary as much as ± 15% depending on the
membrane. Flow rates determined with new hollow fiber final filter installed. Flow rate from final
filter decreases with use.

Weight (dry):

60 lbs. (27.2 kg)

Feedwater Requirements

Type:

Prepurified

via

reverse

osmosis, distillation or

deionization, with a conductivity of < 100µS (Tap
water feed not recommended)

Temperature:

10-30 degrees Centigrade (50-86 degrees

Fahrenheit)

pH:

4-10

Inlet Pressure and Flow:

0-100 psi (0-7 Bar)

providing 2 liters/minute (0.5 gallons/minute) or
better

Deionization Capacity:
(Based on 70% operating efficiency.
See Table under Feed Water Quality
in Installation Section of the manual)

General

Chemistry

models

1373 Grains as CaCO

3


HPLC Analytical Instrument models
915 Grains as CaCO

3


UF Life Science models
1373 Grains as CaCO

3

Advertising
Ce manuel est liée aux produits suivants: