Legal for trade sealing – Ohaus I5S Indicator Manuel d'utilisation

Page 32

Advertising
background image

31

Legal For Trade Sealing

Legal For Trade Sealing

Legal For Trade Sealing

Legal For Trade Sealing

Legal For Trade Sealing

The

I

5S has been tested and found to

comply with the Class III and IIIL require-
ments of NIST Handbook 44.

After the indicator has been tested and
found to comply with applicable regula-
tions by a local Weights and Measures
official, it may be sealed as follows:

1.

Remove the right end cap and make
sure that the menu lockout switch
on the printed circuit board is set to
the OFF position.

2.

Enter the Setup submenu and go to
the LOCSW parameter. Set the
SETUP and CAL LOCSW param-
eters to ON.

Refer to the Menu

Lock-out Switch Setting on page
2-10 for detailed procedures.

3.

Set the menu lockout switch on the
printed circuit board to the ON posi-
tion.

4.

If an audit trail or paper seal will be
used, install the End Cap and tighten
the two 6-32 phillips pan head
screws.

5.

If a lead and wire seal will be used,
replace the 6-32 phillips pan head
screws with the 6-32 hex socket
security screws and washers en-
closed in the plastic bag containing
the accessories.