Auto Zero Tracking

Auto Zero Tracking

Auto Zero Tracking

Auto Zero Tracking

Auto Zero Tracking

Auto Zero Tracking compensates for
changes in zero reading due to temper-
ature changes or build up of small
amounts of debris on the scale base
platform. Select (0.5, 1, or 3 divisions or
OFF), the value of weight change that
the indicator will disregard in order to
maintain a zero reading.

Auto Zero Tracking occurs only when
the display indicates zero regardless of
the load on the scale base.

NOTE

NOTE

NOTE

NOTE

NOTE: If LFT is set ON, the Auto-Zero
Tracking OFF setting is not available.

To access the current setting, press

when Auto-0 is displayed.

1.

To change the displayed setting,

press

.

2.

To accept the displayed setting,

press

.

To proceed to the next parameter (Auto

Off Timer), press

when Auto-0 is

displayed.

To exit the Auto-Zero Tracking submenu
and store your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press

until

END is displayed then press

. The

scale is now in a weighing mode.