Calibration Submenu

Calibration Submenu

Calibration Submenu

Calibration Submenu

Calibration Submenu

The calibration menu is used to perform
calibration.

If you are already in the main menu, you

can repeatedly press

until CAL is

displayed and proceed with step 3 of the
following procedure. If you are not al-
ready in the main menu, start with step 1.

Procedure

Procedure

Procedure

Procedure

Procedure

Before beginning, make sure there is no
load on the scale base and calibration
weight equal to the Calibration Point is
available.

1.

With the indicator OFF, press and

hold

until SETUP is dis-

played, then release it.

2.

Press

twice to display CAL.

3.

When CAL is displayed, press

to begin calibration.

4.

C 0 will be displayed. With no load

on the platform, press

.

5. When

is released, -C- will be

displayed followed by C n indicat-
ing the calibration weight value.

6. Place the weight on the load re-

ceiver and press

to accept

the calibration data.

7. The display will show -C-, return to

the weighing mode and display the
calibration weight value with the
current unit indicator.