Zero – Ohaus I5S Indicator Manuel d'utilisation

Page 16

Advertising
background image

15

Zero

Zero

Zero

Zero

Zero

Zero specifies the percentage of full
capacity load (2% or 100%) that may be

cleared by pressing

.

A 2% Zero setting is used for hop-
per scales or other large scales
where an accidental zero would
lose the current weight. A 2% Zero
is also required for Canadian Le-
gal for Trade applications. Bench
and counter scales normally use
100% Zero setting.

To access the current setting, press

when ZERO is displayed.

1.

To change the displayed setting,

press

.

2.

To accept the displayed setting,

press

.

To proceed to the next parameter (Cali-

bration Unit), press

when ZERO is

displayed.

To exit the Setup submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press

until

END is displayed then press

. The

scale is now in a weighing mode.

Setup and Calibration

2

2

2

2

2