Setup submenu, Legal for trade – Ohaus I5S Indicator Manuel d'utilisation

Page 15

Advertising
background image

14

Setup Submenu

Setup Submenu

Setup Submenu

Setup Submenu

Setup Submenu

1.

With the indicator OFF, press and

hold

until SETUP is dis-

played, then release it.

2.

When SETUP is displayed, press

to enter the submenu. The

first parameter, Legal For Trade, will
be displayed.

Legal For Trade (LFT)

Legal For Trade (LFT)

Legal For Trade (LFT)

Legal For Trade (LFT)

Legal For Trade (LFT)

This parameter is used to configure the

I

5S for Legal for Trade or non Legal for

Trade use. Setting this parameter to ON
automatically sets certain parameters for
legal for trade requirements, disables
others, and limits overall display resolu-
tion to 1:5,000.

To access the current setting (ON or

OFF), press

when LFT is dis-

played.

1.

To change the displayed setting,

press

.

2.

To accept the displayed setting,

press

. To proceed to the

next parameter (Zero), press

when LFT is displayed.

To exit the Setup submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu. To exit main
menu and return to a weighing mode,

repeatedly press

until END is dis-

played then press

. The scale is

now in a weighing mode.

Legal for Trade OFF
Zero
100%

Calibration

Unit lb

Full Scale

Capacity 5000
Graduation Size 1
Calibration Point 5000
Menu Lock-out Switch

Setup ON
Scale ON
Calibration ON
RS232 ON

SETUP MENU

SETUP MENU

SETUP MENU

SETUP MENU

SETUP MENU

FACTORY DEFAULTS

FACTORY DEFAULTS

FACTORY DEFAULTS

FACTORY DEFAULTS

FACTORY DEFAULTS

2

2

2

2

2

Setup and Calibration

Advertising