Graduation Size

Graduation Size

Graduation Size

Graduation Size

Graduation Size

This parameter specifies the increments
in which the display will count.

If Legal For Trade is set to ON, resolu-
tion (Capacity

÷

Graduations) is limited

to a maximum of 5000 divisions. The
graduation selections available are lim-
ited by the capacity entered under Full
Scale Capacity to prevent a combination
which exceeds 5,000 divisions.

If the graduation size selected does not
meet the resolution requirement, the
Setup menu cannot be exited. An error
message is displayed and you are auto-
matically returned to the GRAD param-
eter. A valid graduation value must be
selected.

To access the current setting, press

when GRAD is displayed.

1.

To change the displayed setting,

press

.

2.

To accept the displayed setting,

press

.

To proceed to the next parameter (Cali-

bration Point), press

when GRAD

is displayed.

To exit the Setup submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press

until

END is displayed then press

. The

scale is now in a weighing mode.

Complete List of Graduation

Complete List of Graduation

Complete List of Graduation

Complete List of Graduation

Complete List of Graduation

Selections

Selections

Selections

Selections

Selections

.0005, .001, .002,
.005, .01, .02, .05,
.1, .2, .5, 1

1

1

1

1, 2, 5, 10

2

2

2

2

2