Chapter 6: Maintaining the Protector Echo

Product Service 1-800-522-7658

93

Original instructions

Electro (Acid) Sensor Replacement

(every 2 years)


We recommend the following:

We recommend that you follow all your existing lab safety protocols as

they apply to this procedure.

That you have another person to assist you when possible.

Please thoroughly read through before attempting it.


Items needed:

6ft ladder

Piece of cardboard

Maintenance Card

Small slotted screwdriver and small Phillips screwdriver


**Some GFH models have #2 Phillips screws for everything listed below**

Small Phillips Screwdriver

4mm Allen wrench (for Control module screws)

Torx TX20 (back panel screws)

Torx TX10 (Electro/Acid sensor board)


Overview
1.

Write down all parameters under “Hood Status.”

2.

Turn off hood.

3.

Remove front panel and identify electrical connection whether it is either

hardwired or plugged into a receptacle.

4.

Disconnect all connections from control module that connects to the

technology.

5.

Remove control module & electrical cord set if applicable to your set up.

6.

Snake all remaining wires that are routed through the superstructure to control

module.

7.

Remove back panel to control module.

8.

Disconnect and remove old sensor.

9.

Install new sensor.

10.

Attach back panel to control module.

11.

Snake all remaining wires that are routed through the superstructure to control

module.

12.

Connect the electrical cord set to the controller, if applicable to your set up.

13.

Attach the control module to the mounting bracket.

14.

Connect all remaining wires.

15.

Turn power back on.