Chapter 3: Getting Started

Product Service 1-800-522-7658

14

Original instructions

protective earthing ground lug, no other conductors should be connected to this
grounding lug. Using wire nuts, connect the MAINS supply conductors to the fume
hood supply wires. Ensure that the wires are connected as per the appropriate wire
color codes for the input voltage. For 115V Phase (Hot) is black and Neutral is
white, for 230V Phase1 is brown and Phase2 is blue. Refer to the wiring diagram for
your Protector Echo in Appendix C: Protector Echo Specifications.

Figure 3-2