Chapter 6: Maintaining the Protector Echo
Product Service 1-800-522-7658
91
Original instructions
(You will need this information only if you have gGuard installed.)
If DHCP is
yes
then you have a dynamic IP if
no
then you have a static IP which
you must reenter completely.
Elapsed Time
Filter X D (this is the # of days since install)
Fans
Set point Dflt. XXXX rpm
Set point XXXX rpm
Semicon S. trig value (example 3500)
Trig. X
Delay X
Sash
Which type (example. normal polarity)
Trig X %
Now press
“X”
to return to the previous menu.”
Step 4 – Turn off the hood
Press the fan icon to turn off the controller waiting the 30 seconds to completely
shut down. Also if the light is on, turn it off as well.
Step 5 – Remove failed fan box
Take off the secondary filter that is directly above the fan you are replacing. Once
you have it removed unplug the following:
The left fan power supply wire
Air sampling tube
The right fan power supply wire to the next fan (if applicable)
***Again make note of the
micro
switch settings of that fan.***
CAUTION: Do not contact fan wheel while still in motion.
PRUDENCE: Ne pas contacter roue du ventilateur tout en en
mouvement.
!
- Protecto Ech Filtered Fume Hoods 183 Series 6' Protecto Ech Filtered Fume Hoods 183 Series 5' Protecto Ech Filtered Fume Hoods 183 Series 4 Protecto Ech Filtered Fume Hoods 182 Series 8' Protecto Ech Filtered Fume Hoods 182 Series 6' Protecto Ech Filtered Fume Hoods 182 Series 5' Protecto Ech Filtered Fume Hoods 182 Series 4' Protecto Ech Filtered Fume Hoods 181 Series 8' Protecto Ech Filtered Fume Hoods 181 Series 6' Protecto Ech Filtered Fume Hoods 181 Series 5' Protecto Ech Filtered Fume Hoods 181 Series 4' Protecto Ech Filtered Fume Hoods 180 Series 8' Protecto Ech Filtered Fume Hoods 180 Series 6' Protecto Ech Filtered Fume Hoods 180 Series 5' Protecto Ech Filtered Fume Hoods 180 Series 4'