Detecting a breakthrough – Labconco Protecto Ech Filtered Fume Hoods 183 Series 8' Manuel d'utilisation
Page 65

Chapter 5: Using the Protector Echo
Product Service 1-800-522-7658
59
Original instructions
Figure 5-2
The Modular Neutrodine Filtration Technology
Detecting a Breakthrough
Detecting a breakthrough of the lower primary filter is automatically completed
with a broad sensor, a target acid gas sensor, and a lab ambient air sensor.
Broad Sensor (Solvents)
The first sensor is based on a semi-conductor metal oxide sensing chip with
integrated heater. In the presence of detectable gas, sensor concentration
increases with electrical output signal which corresponds to the gas concentration.
The first sensor detects a large range of chemicals, such as solvents, but is not
sensitive to inorganic acids and some halogenated chemicals.
Safety
Backup
Filter
Fan
Primary
Filter
Light and
Pre-Filter
- Protecto Ech Filtered Fume Hoods 183 Series 6' Protecto Ech Filtered Fume Hoods 183 Series 5' Protecto Ech Filtered Fume Hoods 183 Series 4 Protecto Ech Filtered Fume Hoods 182 Series 8' Protecto Ech Filtered Fume Hoods 182 Series 6' Protecto Ech Filtered Fume Hoods 182 Series 5' Protecto Ech Filtered Fume Hoods 182 Series 4' Protecto Ech Filtered Fume Hoods 181 Series 8' Protecto Ech Filtered Fume Hoods 181 Series 6' Protecto Ech Filtered Fume Hoods 181 Series 5' Protecto Ech Filtered Fume Hoods 181 Series 4' Protecto Ech Filtered Fume Hoods 180 Series 8' Protecto Ech Filtered Fume Hoods 180 Series 6' Protecto Ech Filtered Fume Hoods 180 Series 5' Protecto Ech Filtered Fume Hoods 180 Series 4'