Chapter 3: Getting Started
Product Service 1-800-522-7658
40
Original instructions
You should see this menu again:
>Hood integration
Fan set point
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash sensor trigger
Software releases
Audible alarm Mgmt.
Arrow down to
“Electro S. alarm Delay”
press
“Ok”
to select it.
You will see:
05 Minutes
1 Min
60 Max
Make the
0
a
1
for
15 Minutes
Press
“Ok”
then
“Confirm new value”
press
“Ok”
to confirm.
You should see this menu again:
>Hood integration
Fan set point
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash sensor trigger
Software releases
Audible alarm Mgmt.
Arrow down to
“Sash sensor trig.”
Press
“Ok”
to select it.
Refer back to your notes from step 2 for the type of
Sash sensor
and
trig. X%
value
Arrow down to your sash then press
“Ok”
to select it.
“Confirm new value”
press
“Ok.”
You will see
Sash sensor trigger.
Enter the
trig%
from step 2.
“Confirm new value”
press
“Ok.”
For example type of sash
“inverted polarity” trig. 65%
Press
“X”
to return to the previous menu.
- Protecto Ech Filtered Fume Hoods 183 Series 6' Protecto Ech Filtered Fume Hoods 183 Series 5' Protecto Ech Filtered Fume Hoods 183 Series 4 Protecto Ech Filtered Fume Hoods 182 Series 8' Protecto Ech Filtered Fume Hoods 182 Series 6' Protecto Ech Filtered Fume Hoods 182 Series 5' Protecto Ech Filtered Fume Hoods 182 Series 4' Protecto Ech Filtered Fume Hoods 181 Series 8' Protecto Ech Filtered Fume Hoods 181 Series 6' Protecto Ech Filtered Fume Hoods 181 Series 5' Protecto Ech Filtered Fume Hoods 181 Series 4' Protecto Ech Filtered Fume Hoods 180 Series 8' Protecto Ech Filtered Fume Hoods 180 Series 6' Protecto Ech Filtered Fume Hoods 180 Series 5' Protecto Ech Filtered Fume Hoods 180 Series 4'