Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity Manuel d'utilisation
Page 55
Advertising

45
IM 12D06D05-01F-E
9
QUALITY
INSPECTION
3/3
QIS 12D06D05-01E
Table 4
Data Display
Current Output (mA DC)
4
4 ±0.02
12
12 ±0.02
20
20 ±0.02
After all tests are completed,
a. Touch the [Exit] twice to return to the “Service” display.
b. Select “Normal” in “Key”.
c. Touch the [Home] icon to return to the initial display.
Figure 1 Testing Circuit and Test Equipment