IM 12B6C3-F-E

11-1 Test Certificate

Test

EXA Series

Certificate

Model PH202

Transmitter for pH / ORP

1.

Introduction

This inspection procedure applies to the model PH202 converter. There is a serial number, unique to
the instrument, which is stored in non-volatile memory. Each time the converter is powered up, the
serial number is shown in the display. An example is shown below, for details see the Users manual:

Unique Number

Line Number

ATE (automatic test equipment no.)

Month code

Year code

2.

General Inspection

Final testing begins with a visual inspection of the unit to ensure that all the relevant parts are present
and correctly fitted.

3.

Safety Test

The (-) minus and the external ground terminal of the housing are connected to a Voltage generator
(100 VDC). The measured impedance value should be over 9.5 MΩ.

Terminal 12 and the external ground terminal of the housing are connected to a Voltage generator
(500 VAC RMS) for 1 minute. The leakage current should remain below 8 mA.

4.1 Accuracy Testing

Our automated testing facility checks the accuracy of the dual high inputs of the instrument using a
calibrated variable resistor (decade resistor box) to simulate sensor mV’s.

4.2 Accuracy Testing of all supported temperature elements

Our automated testing facility checks the input accuracy of the instrument using a calibrated variable
resistor (decade resistor box) to simulate the resistance of all temperature elements.

11. Test Certificate