Chapter 3: Getting Started

Product Service 1-800-522-7658

16

Original instructions

The fluorescent light has been mounted behind the front panel and is sealed from
vapors inside the hood structure. The fluorescent ballast is located in a separate
box on top of the fume hood adjacent to the main junction box. To change the
fluorescent light bulbs in your hood, you must first remove the front panel(s) and
then the cover plate(s), which is secured with six (6) acorn nuts (See Fig. 3-3).
Twist and remove any deflective bulbs. Reverse order to reassemble.

Connecting the Service Lines to the

Protector Fume Hood

The hoods with service fixtures have been plumbed from the valve to the hose
connector or gooseneck for your installation convenience. Supply tubing shall be
provided by the qualified installer. Tubing can enter the hood from above,
through the back, or through the work surface to make these connections to the
service fixtures.


NOTE: Inspect all fittings for leakage. Tighten the fittings
slightly if needed.

NOTE : Inspecter toutes les installations à la recherche de
fuite. Resserrer les installations légèrement si nécessaires.

CAUTION: Do not use oxygen with any standard service
fixture. Contact Labconco Customer Service for oxygen
fixture information.

PRUDENCE : Ne pas utiliser de l'oxygène avec l'accessoire de
service standard. Contacter le Service Clientèle de Labconco
pour les informations d'accessoire d'oxygène.

Should access to the hood plumbing fixture bodies be required, remove the
service access plate on the hood front corner posts by loosening their individual
screws (see item 11, Figure A-1 in Appendix A). The valve body will now be
fully exposed for any service work that may be necessary. The service fixtures
supplied on your laboratory hood are designed for use with the following services:

Air

Hot Water

Vacuum

Cold Water

Natural Gas – See caution below

WARNING: Contact Labconco Customer Service directly
before using any service other than those listed above in these
valves to assure full compatibility.

AVERTISSEMENT : Contacter le Service Clientèle de
Labconco directement avant d'utiliser n'importe quel service
autre que ceux énumérés au-dessus dans ces soupapes pour
assurer une pleine compatibilité.

!

!

!