Environmental Conditions
• Indoor use only.
• Maximum altitude: 10,000 feet (3,048 meters).
• Ambient temperature range: 41° to 104°F (5° to 40°C).
• Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
• Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
• Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
• Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.
• Electrical Ratings
Volts AC
Phase
Cycle
AMP*
115 1 50/60 10
115 1
50
10
115 1
60
10
230 1 50/60 5
230 1
50
5
230 1
60
5
*Does not include current rating of receptacles.