Chapter 2: Prerequisites

Product Service 1-800-522-7658

8

Original instructions

Electrical Requirements

The Protector Hood models feature internal wiring for the fluorescent light
assembly and light switch. All internal wiring is terminated at the single point
wiring junction box for hook-up by a qualified electrician. The blower switch,
and light switch wires are also terminated at the single point wiring junction box
for hook-up by a qualified electrician. Refer to Chapter 3: Getting Started and
Appendix C: Protector XL Specifications for the wiring diagram for proper
electrical installation.

Les modèles Protège-capot disposent câblage interne de l'assemblage de lumière
fluorescente et interrupteur de lumière. Tout le câblage interne est terminé à la
boîte de jonction point de câblage unique pour le raccordement par un électricien
qualifié. Le bouton du ventilateur, et fils de l'interrupteur de lumière sont
également mis fin à la boîte de jonction point de câblage unique pour le
raccordement par un électricien qualifié. Reportez-vous au Chapitre 3: Mise en
route et à l'Annexe C: Spécifications Premier Protector pour le schéma de câblage
pour l'installation électrique correcte.

Service Line Requirements

All service lines to the laboratory fume hood should be ¼ inch outside diameter,
copper (brass for natural gas), and equipped with an easily accessible shut-off
valve, should disconnection be required. If the service line pressure exceeds 40
PSI, it must be equipped with a pressure regulator to reduce the line pressure.
Please check with local codes for other requirements.

Space Requirements

The dimensions for the different models are shown in Appendix B: Protector XL
Dimensions
.