Chapter 2: prerequisites, Original instructions – Labconco Protector XL Laboratory Hoods 1118 Series Manuel d'utilisation

Page 11

Advertising
background image

Chapter 2: Prerequisites

Product Service 1-800-522-7658

7

Original instructions

Proper blower selection can be determined from these exhaust requirements and
the total system static pressure loss. Contact Labconco Customer Service for
assistance in sizing a blower system.

Airflows and Static Pressure

Eco Energy Saving Air Foil

Flush Foil

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Airflow Volumetric Rate (CFM) @

Sash at

Sash at 62.5% Static Pressure (inches of water)

Static Pressure (inches of water)

Full open

open

4'

5'

6'

8'

4'

5'

6'

8'

at 28"

at 18"

Hood

Hood

Hood

Hood

Hood

Hood

Hood

Hood

XL

100

160

725, 0.27"

955, 0.34" 1180, 0.46" 1640, 0.31"

745, 0.28"

985, 0.36" 1220, 0.49" 1695, 0.33"

80

128

580, 0.17"

765, 0.22" 945, 0.29" 1310, 0.20"

595, 0.18"

790, 0.23" 975, 0.31" 1355, 0.21"

60

96

435, 0.10"

575, 0.12" 710, 0.17"

985, 0.11"

445, 0.10"

590, 0.13" 730, 0.18" 1015, 0.12"

50

80

365, 0.07"

480, 0.09" 590, 0.11"

820, 0.08"

375, 0.07"

495, 0.09" 610, 0.12" 850, 0.08"

40

64

290, 0.04"

380, 0.05" 470, 0.07"

655, 0.05"

300, 0.05"

395, 0.06" 490, 0.08" 680, 0.05"

Eco Energy Saving Air Foil

Flush Foil

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Airflow Volumetric Rate (CFM) @

Sash at

Sash at

Static Pressure (inches of water)

Static Pressure (inches of water)

62.50%

Full open

4'

5'

6'

8'

4'

5'

6'

8'

open at 18"

at 28"

Hood

Hood

Hood

Hood

Hood

Hood

Hood

Hood

XL

100

62

450, 0.11"

595, 0.13" 735, 0.18" 1025, 0.12"

465, 0.11"

615, 0.14" 765, 0.19" 1060, 0.13"

80

50

365, 0.07"

480, 0.09" 590, 0.11"

820, 0.08"

375, 0.07"

495, 0.08" 610, 0.11" 850, 0.08"

60

38

270, 0.04"

360, 0.05" 440, 0.07"

615, 0.04"

280, 0.04"

370, 0.05" 460, 0.07" 635, 0.05"

Airflows and Static Pressure

Eco Energy Saving Air Foil

Flush Foil

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Airflow Volumetric Rate (CFM) @

Sash at

Sash at 62.5% Static Pressure (inches of water)

Static Pressure (inches of water)

Full open

open

4'

5'

6'

8'

4'

5'

6'

8'

at 28"

at 18"

Hood

Hood

Hood

Hood

Hood

Hood

Hood

Hood

XL

100

160

705, 0.25"

930, 0.32" 1150, 0.44" 1600, 0.30"

725, 0.26"

960, 0.34" 1190, 0.47" 1655, 0.32"

W/Bypass

80

128

565, 0.16"

745, 0.21" 920, 0.28" 1280, 0.19"

580, 0.17"

770, 0.22"

950, 0.30" 1325, 0.21"

Block

60

96

425, 0.09"

560, 0.12" 690, 0.16"

960, 0.11"

435, 0.09"

575, 0.12"

715, 0.17" 995, 0.12"

50

80

350, 0.06"

465, 0.08" 575, 0.11"

800, 0.08"

365, 0.07"

480, 0.09"

595, 0.12" 830, 0.08"

40

64

280, 0.04"

370, 0.05" 460, 0.07"

640, 0.05"

290, 0.04"

385, 0.05"

475, 0.08" 660, 0.05"

Eco Energy Saving Air Foil

Flush Foil

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Airflow Volumetric Rate (CFM) @

Sash at

Sash at

Static Pressure (inches of water)

Static Pressure (inches of water)

62.50%

Full open

4'

5'

6'

8'

4'

5'

6'

8'

open at 18"

at 28"

Hood

Hood

Hood

Hood

Hood

Hood

Hood

Hood

XL

100

62

440, 0.10"

580, 0.12" 720, 0.17" 1000, 0.12"

455, 0.10"

600, 0.13"

745, 0.18" 1035, 0.12"

W/Bypass

80

50

350, 0.06"

465, 0.08" 575, 0.11"

800, 0.08"

365, 0.07"

480, 0.09"

595, 0.12" 830, 0.08"

Block

60

38

265, 0.04"

350, 0.05" 430, 0.06"

600, 0.04"

275, 0.04"

360, 0.05"

445, 0.07" 620, 0.05"

Airflows and Static Pressure

Eco Energy Saving Air Foil

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Sash at

Sash at 62.5% Static Pressure (inches of water)

Full open

open

10'

12'

16'

at 28"

at 18"

Hood

Hood

Hood

XL

100

160

2100, 0.45" 2560, 0.62" 3500, 0.37"

W/Bypass

80

128

1680, 0.28" 2050, 0.39" 2800, 0.23"

Block

60

96

1260, 0.16" 1535, 0.22" 2100, 0.13"

50

80

1050, 0.11" 1280, 0.16" 1750, 0.09"

40

64

840, 0.07" 1025, 0.10" 1400, 0.06"

Eco Energy Saving Air Foil

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Sash at

Sash at

Static Pressure (inches of water)

62.50%

Full open

10'

12'

16'

open at 18"

at 28"

Hood

Hood

Hood

XL

100

62

1300, 0.17" 1585, 0.24" 2170, 0.14"

W/Bypass

80

50

1050, 0.11" 1280, 0.16" 1750, 0.09"

Block

60

38

800, 0.06"

970, 0.09" 1330, 0.05"

Advertising