Jenway Genova Nano French Manuel d'utilisation

Page 16

Advertising
background image

14

18

The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

5. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

6. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

9. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

10. The Calibration Solution Measurement icon will

be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

18

The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

5. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

6. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

9. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

10. The Calibration Solution Measurement icon will

be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

18

The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

5. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

6. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

9. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

10. The Calibration Solution Measurement icon will

be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

3. L'icône de mesure de solution à blanc s'affiche.

Ouvrez le couvercle de l'instrument puis, à
l'aide d'une pipette, placez une aliquote de
2,0 μl de la solution à blanc de matrice sur la
tête de lecture inférieure.

Fermez le couvercle de l'instrument et appuyez
sur la touche située près de l'icône de validation
pour lancer la mesure à blanc.

4. L'appareil produit cinq relevés à 260 et

330 nm, à la fois pour les trajets optiques 0,5
et 0,2 mm.

6. L'icône de mesure de la solution d'étalonnage

s'affiche. Ouvrez le couvercle de l'instrument
et utilisez un chiffon non-pelucheux pour
nettoyer les têtes de lecture.

7. À l'aide d'une pipette, placez une aliquote

de 2,0 μl de solution d'étalonnage sur la tête
de lecture inférieure. Fermez le couvercle de
l'instrument et appuyez sur la touche située
près de l'icône de validation pour lancer la
mesure de la solution d'étalonnage.

8. L'appareil produit cinq relevés à 260 et

330 nm, à la fois pour les trajets optiques 0,5
et 0,2 mm.

10. L'icône de mesure de la solution d'étalonnage

s'affiche pendant une seconde. Ouvrez le
couvercle de l'instrument et utilisez un chiffon
non-pelucheux pour nettoyer les têtes de
lecture.

11. À l'aide d'une pipette, placez une aliquote

de 2,0 μl de solution d'étalonnage sur la tête
de lecture inférieure. Fermez le couvercle de
l'instrument et appuyez sur la touche située
près de l'icône de validation pour lancer la
mesure de la solution d'étalonnage.

5. Si les valeurs mesurées sont comprises dans les tolérances requises, l'icône de validation de test

s'affiche. Appuyez sur la touche située près l'icône de validation pour continuer ou sur le bouton
situé près de l'icône de croix pour annuler le processus d'étalonnage.

9. Si les valeurs mesurées sont comprises dans les tolérances requises, l'icône de validation de test

s'affiche. Appuyez sur la touche située près l'icône de validation pour continuer ou sur le bouton
situé près de l'icône de croix pour annuler le processus d'étalonnage.

12. L'appareil produit cinq relevés à 260 et 330 nm, avec à la fois les les trajets optiques 0,5

et 0,2 mm.

13. Si les valeurs mesurées sont comprises dans les tolérances requises, l'icône de validation de test

s'affiche et l'instrument calcule les nouveaux facteurs d'étalonnage de trajet optique. Appuyez sur
la touche située près de l'icône de validation pour continuer ou sur la touche située près de l'icône
de croix pour annuler le processus d'étalonnage.

14. Les nouvelles valeurs de facteur d'étalonnage doivent maintenant être vérifiées en mesurant de

nouveau les valeurs air, blanc et étalon.