4. Sélectionnez le chiffre à modifier à l'aide des touches situées dans la partie inférieure de l'écran.

Utilisez les touches situées près des icônes de flèche pour augmenter ou diminuer ce chiffre. Appuyez
sur la touche située près de l'icône de validation pour enregistrer les modifications.

5.2.3

étalonnage et vérification de l'accessoire

Si un message d'erreur s'affiche pendant la procédure d'étalonnage et de vérification de
l'accessoire, reportez-vous à la Section 9 pour plus de détails.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

2. Sélectionnez le chiffre à modifier à l'aide des

touches situées dans la partie inférieure de
l'écran. Utilisez les touches situées près des
icônes de flèche pour augmenter ou diminuer le
chiffre.

Appuyez sur la touche située près de l'icône de
validation pour enregistrer les modifications.

3. Saisissez le trajet optique déclaré (1,0 mm pour le

kit de solution-étalon d'étalonnage fourni) pour
les valeurs de solution-étalon ADN, en mm, dans
les paramètres de l'accessoire en appuyant sur la
touche située près de l'icône de valeur de trajet
optique de la norme d'étalonnage.

1. Appuyez sur la touche située près de l'icône

d'étalonnage de trajet optique pour lancer la
séquence d'étalonnage de l'accessoire.

2. L'icône de mesure d'air s'affiche. Utilisez un

chiffon non-pelucheux pour nettoyer les têtes
de lecture supérieure et inférieure, fermez le
couvercle de l'instrument et appuyez sur la
touche située près de l'icône de validation pour
continuer. Appuyez sur la touche située près de
l'icône de croix pour revenir à l'écran du menu
des paramètres de l'accessoire à micro-volume.

L'appareil produit cinq relevés clairs et obscurs à
260 et 330 nm, avec à la fois les trajets optiques
0,5 et 0,2 mm.