Maintenance and Troubleshooting

35

5

Display

Action

Your blood glucose may be extremely high or a meter or a test strip error has

occurred.

• If your test result matches how you feel, contact your healthcare professional

immediately.

• If your test result does not match how you feel, repeat the blood glucose test.

See Chapter 2, Unusual Blood Glucose Results.

• If the E-3 code still appears for your blood glucose test, your blood glucose

result may be extremely high and above the system’s reading range.

Contact your healthcare professional immediately.

• If the second test result does not match how you feel, perform a control

test with the control solution and a new test strip.

• If the control result is within the acceptable range, review the proper

testing procedure and repeat the blood glucose test with a new test

strip.

• If the control result is not within the acceptable range, see Chapter 4,

Understanding Out-of-Range Control Results.

Not enough blood or control solution was drawn into the test strip for

measurement or was applied after the test had started. Discard the test strip and

repeat the blood glucose or control test.
This message may appear when using a white code chip in the meter. It means

the white code chip is from an expired lot of test strips. Turn the meter off and

insert a black activation chip. Make sure the time and date in the meter are

correct.
Blood or control solution was applied to the test strip before the flashing drop

symbol appeared on the display. Discard the test strip and repeat the blood

glucose or control test.
An electronic error occurred, or in rare cases, a used test strip was removed and

reinserted. Turn the meter off and on, or take the battery out for 20 seconds and

reinsert it. Perform a blood glucose or control test.