Control Tests

28

4

Touch the drop to the

front edge of the yellow

window of the test strip

until you see flash.
Do not put control

solution on top of the

test strip.

5

Remove the cap from the

control solution bottle.

Wipe the tip of the bottle

with a tissue.
Squeeze the bottle until a

tiny drop forms at the tip.

4

Insert the test strip into

the meter in the direction

of the arrows.
Place the meter on a flat

surface.

Check the use by date

on the test strip

container. Do not use

test strips past the

use by date.

1

Performing a Control Test

You need the meter with pre-inserted activation chip, a test strip, and control solution Level 1, Level 2, or

both. The control level is printed on the bottle label.

Select the control

solution to test. You will

enter the level later in

the test.

3

2

or

6

When you see flash,

there is enough control

solution in the test strip.

Wipe the tip of the bottle

with a tissue. Cap the

bottle tightly.