Control Tests

27

4

When to Perform a Control Test

Performing a control test lets you know the meter and test strips are working properly. You should

perform a control test when:

• you open a new test strip box.

• you left the test strip container open.

• you think the test strips are damaged.

• you want to check the meter and test strips.

• the test strips were stored in extreme temperatures, humidity, or both.

• you dropped the meter.

• your test result does not match how you feel.

• you want to check if you are performing the test correctly.

Chapter 4: Control Tests

About the Control Solution

• Use only Accu-Chek Aviva control solution.

• Close the control solution bottle tightly after use.

• Write the date you open the control solution bottle on the bottle label. The control solution must be

discarded 3 months from the date the control solution bottle was opened (discard date) or on the use by

date on the bottle label, whichever comes first.

• Do not use control solution that is past the use by or discard date.

• Refer to the control solution package insert for control solution storage conditions.

• The meter automatically recognizes the difference between the control solution and blood.

• The control results are not displayed in memory.

• The control solution can stain fabric. Remove stains by washing with soap and water.