Maintenance and Troubleshooting

5

Display

Action

An electronic error occurred, or in rare cases, a used test strip was removed and

reinserted. Turn the meter off and on, or take the battery out for 20 seconds and

reinsert it. Perform a blood glucose or control test.
The temperature is above or below the proper range for the system. Refer to the

test strip package insert for system operating conditions. Move to an area with the

appropriate conditions, wait 5 minutes, and repeat the blood glucose or control test.

Do not artificially heat or cool the meter.
The battery is almost out of power. Change the battery now. If the message

reappears after the battery has been replaced, remove the battery again, press any

meter button, then reinsert the battery.
The time and date settings may be incorrect. Make sure the time and date are

correct and adjust, if necessary.