Maintenance and troubleshooting – Accu-Chek Aviva Plus black Manuel d'utilisation

Page 39

Advertising
background image

37

Maintenance and Troubleshooting

5

Display

Action

A pre-meal marker was assigned to this test result.

A post-meal marker was assigned to this test result.

A pre-meal marker was assigned to this test result and the post-meal test reminder

has been activated.
The test strip may be damaged or not properly inserted. Remove and reinsert the

test strip, or replace it if damaged.

Your blood glucose may be extremely high or a meter or a test strip error has

occurred.

• If your test result matches how you feel, contact your healthcare professional

immediately.

• If your test result does not match how you feel, repeat the blood glucose test. See

Chapter 2, Unusual Blood Glucose Results.

• If the E-3 code still appears for your blood glucose test, your blood glucose

result may be extremely high and above the system’s reading range.

Contact

your healthcare professional immediately.

• If the second test result does not match how you feel, perform a control test

with the control solution and a new test strip.

• If the control result is within the acceptable range, review the proper

testing procedure and repeat the blood glucose test with a new test

strip.

• If the control result is not within the acceptable range, see Chapter 4,

Understanding Out-of-Range Control Results.

Not enough blood or control solution was drawn into the test strip for measurement

or was applied after the test had started. Discard the test strip and repeat the blood

glucose or control test.
Blood or control solution was applied to the test strip before the flashing drop

symbol appeared on the display. Discard the test strip and repeat the blood glucose

or control test.