Maintenance and Troubleshooting
5
What to Clean and Disinfect
The following parts of the meter should be cleaned and disinfected:
• The area around slots and openings (do not get any moisture in slots or openings)
• The meter display
• The entire meter surface
How to Clean and Disinfect the Meter
w
WARNING
Failure to follow these instructions will damage the meter and stop it from working properly.
• DO NOT clean or disinfect the meter while performing a blood glucose or control test.
• DO NOT get any moisture in slots or openings.
• DO NOT spray anything onto the meter.
• DO NOT immerse the meter in liquid.
• Always use the same product for both cleaning and disinfecting.
Wash hands thoroughly
with soap and water.
1
A separate Sani-Cloth
Plus should be used for
cleaning and disinfection.
For disinfecting the meter,
get a new cloth and
repeat step 2, making
sure that the surface
stays wet for 3 minutes.
Make sure that no
solution is seen in any
slot or opening. Ensure
the meter is completely
dry before performing a
blood glucose test.
3
Wash hands thoroughly
with soap and water.
Turn the meter off and
wipe the entire meter
surface with an approved
cleaning and disinfecting
product. Carefully wipe
around the test strip slot
and other openings.
Make sure that no liquid
enters any slot or
opening.
2
4
52596_07022433002_EN_CA.indb 35