Control Tests

4

Understanding Out-of-Range Control Results

w

WARNING

The control ranges are printed on the test strip container label. If the control result is out of range,

check this list to help solve the problem.

Troubleshooting Checks

Action

1. Were the test strips or

control solutions

expired?

Discard the test strips or control solution if either is

past the use by date. If the control solution was opened

more than 3 months ago, discard it. Repeat the control

test with an unexpired test strip and an unexpired

control solution.

2. Did you wipe the tip of the control

solution bottle before use?

Wipe the tip of the bottle with a tissue. Repeat the

control test with a new test strip and a fresh drop of

control solution.

3. Were the caps on the test strip container

and the control solution bottle always

closed tightly?

Replace the test strips or control solution if you think

either was uncapped for some time. Repeat the control

test.

4. Was the test strip used immediately after

it was removed from the test strip

container?

Repeat the control test with a new test strip and a fresh

drop of control solution.

5. Were the test strips and control solutions

stored in a cool, dry place?

Repeat the control test with a properly stored test strip

or control solution.

6. Did you follow the directions?

Read Chapter 4, Control Tests, and repeat the control

test.

7. Did you choose the correct control

solution level, either 1 or 2, when you

performed the control test?

If you chose the wrong control solution level, you can

still compare the control result to the range printed on

the test strip container.

8. Are you still unsure of the problem?

Contact Accu-Chek Customer Care at 1-800-363-7949.