Control tests – Accu-Chek Aviva Plus black Manuel d'utilisation

Page 32

Advertising
background image

30

Control Tests

4

Performing a Control Test

You need the meter, a test strip, and control solution Level 1 or Level 2.

1

Check the use by date on

the test strip container.

Do not use test strips
past the use by date.

2

Insert the test strip into

the meter in the direction

of the arrows.
Place the meter on a flat

surface.

3

Select the control solution

to test. You will enter the

level later in the test.

4

Remove the cap from the

control solution bottle.

Wipe the tip of the bottle

with a tissue.
Squeeze the bottle until a

tiny drop forms at the tip.

5

Touch the drop to the

front edge of the yellow

window of the test strip

until you see flash.
Do not put control

solution on top of the

test strip.

When you see flash,

there is enough control

solution in the test strip.

6

Wipe the tip of the bottle

with a tissue. Cap the

bottle tightly.

or

52596_07022433002_EN_CA.indb 30

4/29/14 9:20 AM

Advertising