Control Tests

4

7

The control result, the

bottle symbol, and a

flashing

L appear on the

display. Do not remove

the test strip yet.
Press to mark the

control result as a

Level 1. Press a

second time to mark the

control result as a

Level 2.

8

Press and release to set the control level in the

meter.
OK and the control result alternate on the display if

the control result is in range.
Err and the control result alternate on the display if

the control result is not in range.
Remove and discard the used test strip.
After a successful test, the meter turns off 5 seconds

after the test strip is removed.

(for example)

(for example)