Metrohm 756 KF Coulometer Manuel d'utilisation

Page 88

Advertising
background image

4.2 Remote control commands

756/831 KF Coulometer, Instructions for Use

80

Monitoring of reagent live time.

.Determ: Number of determinations.

.DCounter: Counter of determinations already carried out.

.MaxTime: Maximum live time of reagent in days.

.TCounter: Time already elapsed since last reagent change.

.ReagCap: Reagent capacity in mg water.

.RCounter: Counter of reagent capacity.

.ClearCount: Clears all above counters.

.Drift:

Stable drift in ug/min.

4.2.2.32. Config.Monitoring.Change $G,

$S

Config.Monitoring.Change.Status auto,

man.,

OFF

Config.Monitoring.Change.WaitTime

0...999 999

Config.Monitoring.Change.AspVol 0...100...9999

Config.Monitoring.Change.SolventVol 0...100...9999

Config.Monitoring.Change.Rinse

0...9999

Config.Monitoring.Change.NoRinse

1...9

Changing of reagent. With a connected Dosino, the reagent is changed with

&Config.Monitoring.Change $G. The nod &Config.Monitoring.change.Status has to
be

OFF.

Parameters for automatic reagent change:

.WaitTime: Waiting time in s after switching off the stirrer.

.AspVol: Volume in ml of used reagent to be aspirated.

.SolventVol: Volume in ml of new reagent to be added.

.Rinse:

Volume in ml of rinsing reagent.

.NoRinse: Number of rinsing cycles.

4.2.2.33. Config.Monitoring.Validation.Status ON,

OFF

Config.Monitoring.Validation.Interval 1...365...9999

Config.Monitoring.Validation.Counter

0...9999

Config.Monitoring.Validation.ClearCount $G

Monitoring of validation.

.Interval: Time interval in days for validation.

.Counter: Time counter in days since last validation.

.ClearCount: Clears the above counter.

4.2.2.34. Config.Monitoring.Service.Status ON,

OFF

Config.Monitoring.Service.Date XXXX-XX-XX

Monitoring of service interval.

4.2.2.35. Config.Monitoring.DiagRep ON,

OFF

Printing of system test report after each switching on of the Coulometer.

4.2.2.36. Config.PeriphUnit.CharSet1

Epson, Seiko, Citizen, HP, IBM

Config.PeriphUnit.CharSet2

Selection of the character set and the graphics control characters for COM1 resp.

COM2 of the Coulometer.

Advertising