4.2 Remote control commands

756/831 KF Coulometer, Instructions for Use

92

Setup.Lock.Config ON, OFF

Setup.Lock.Parameter ON, OFF

Setup.Lock.SmplData ON, OFF

Setup.Lock.UserMeth.Recall ON, OFF

Setup.Lock.UserMeth.Store ON, OFF

Setup.Lock.UserMeth.Delete ON, OFF

Setup.Lock.Exchange ON, OFF

Setup.Lock.Display ON, OFF

ON means disable the corresponding function:

.Keyboard: Disable all keys of the Coulometer

.Config:

Disable the <CONFIG> key

.Parameter: Disable the <PARAM> key

.SmplData:

Disable the <SMPL DATA> key

.UserMeth.Recall: Disable "recall" in <USER METH> key

.UserMeth.Store: Disable "store" in < USER METH > key

.UserMeth.Delete: Disable "delete" in < USER METH > key

.Exchange:

Disable the <EXCH> key

.Display:

Disable the display, i.e. it will not be written to by the device program

of the Coulometer and can be operated from the computer.

4.2.2.90. Setup.Mode.StartWait ON, OFF

Setup.Mode.FinWait ON, OFF

Holding points in the method sequence. If they are "ON", the sequence stops until

"OFF" is sent. Switching the instrument on sets both nodes to OFF:

.StartWait: Holding point right after starting a method (holding point after AutoInfo

!".T.GC").

.FinWait: Holding point at the end a method (holding point after AutoInfo !".T.F").

4.2.2.91. Setup.SendMeas.SendStatus ON, OFF

Setup.SendMeas.Interval 0.4...4...16200, MPList

.SendStatus: ON means the automatic transmission of measured values (see

4.2.2.94 and 4.2.2.95) in the inputted interval is active.

.Interval: Time interval (in s) for the automatic transmission of associated meas-

ured values defined under points 4.2.2.95 and 4.2.2.96. The inputted

value is rounded off to a multiple of 0.4. The smallest possible time in-

terval depends on the number of measured values which have to be

sent, on the baud rate, on the load on the interface and on the type of

device connection. With "MPList" the measured values are sent at the

time of their entry into the measured point list.

The automatic transmission is switched on/off with 'SendStatus'.

4.2.2.92. Setup.SendMeas.Select Assembly,

Titrator

Selection of the unit of which the measured values should be sent (4.2.2.95 or

4.2.2.96).

4.2.2.93. Setup.SendMeas.Assembly.CyclNo ON, OFF

Setup.SendMeas.Assembly.I ON, OFF

Setup.SendMeas.Assembly.Meas ON, OFF