15 Refit the cabinet end panel with all

six screws and washers and the

cabinet roof with all six counter

sunk screws. Replace the perspex

viewing screen.
You have now assembled your

Aquatron Water Still. Move the

assembled A4000 to its working

location - using the mounting

bracket, item 7 if the unit is to be

wall mounted. Ensure there is a

space of at least 50cm to the right

of the unit to allow access to the

heating element.

16 Connect the cold water supply to

the mains water inlet, item 18, fig.

1. Note the tubing selected should

have a safe working pressure of at

least equal to the pressure of the

water supply and should be

adequately secured by hose clips.

DO NOT TURN ON WATER
SUPPLY.

17 Electrical Installation

THIS EQUIPMENT MUST BE
EARTHED!

The electrical installation should

only be carried out by a qualified

electrician.
The A4000 is classed as

permanently connected equipment.
The equipment is supplied with

1.5m of flexible, triple core, circular

cable to the following

specification:- 1.5mm

2

, to BS 6500

or equivalent and <HAR> or BASEC

approved.
Connection to the mains electrical

supply, should be via a double pole

isolation switch with a continuous

current carrying capacity of 15A at

250V and overcurrent protection

should be provided by a double

pole approved fuse rated at 15A.
These devices should be sited near

to the equipment and clearly

marked – “Disconnect device for

Aquatron Water Still“
The colour code for the mains

cable is:–
Brown

– Live

Blue

– Neutral

Yellow/Green

– Earth

Mains Cable Replacement

If the mains cable requires

replacement, only specially

prepared spare mains lead obtained

from Bibby Sterilin should be used.

DO NOT SWITCH ON
ELECTRICITY SUPPLY.

18 Connect the reservoir level control

WLS as detailed on page 14 of the

manual.

19 Refer to page 5 for operating

instructions.