[ 16 ]

– the Reservoir pressure switch is

fitted to the distillate collection
vessel.

– the dummy shorting plug has

been removed from it’s socket
and replaced with the lead from
the WCK/N.

3 Switch on the electricity supply –

the green indicator light will
illuminate.

4 Press the ON/OFF switch to the ON

position – the switch will illuminate.

5 Recheck that the CLEAN switch is

in the CLEAN position and
illuminated.

6 Turn on the mains water supply

and using the needle valve on the
side of the cabinet, adjust the flow
rate to approximately 1 Litre/Min.
for A4000 and 2 Litres/Min. for
A8000 and A4000D. Ensure that
the mains water flows to drain
without any restriction.

7 Turn on the supply of deionised or

pre-treated water. Observe that
deionised water enters the boiler
via the float system of the WCK/N.

8 Allow deionised water to flow into

the boiler until the heating
element(s) is covered by
approximately 10mm. At this level
the flow of deionised water into
the boiler should be automatically
cut off by the WCK/N float system.

9 Switch off the CLEAN button by

pressing it for a second time – the
switch light goes out.

The heater(s) will start to warm up
and eventually run at a red glow. (If
heater fails to glow, check that the
flow of mains water is adequate).

10 Allow distillation to commence and

observe that the WCK/N float
system allows fresh deionised water
to enter the boiler and maintain a
satisfactory operating level.

If the deionised supply fails then
the WCK/N will switch off the
heater until it is restored.

11 Replace the perspex viewing

screen.

When any of the Aquatron stills have

been used to produce distilled water

directly from a mains water supply, there

will inevitably be a build-up of scale in

the boiler and on the heating element(s).

To obtain optimum performance from

the still the scale should be r emoved on

a regular basis. The time span between

cleaning depends entirely upon the

hardness of the water supply. In very

hard water areas it may be necessary to

clean the still once a week, wher eas in

soft water areas several weeks may

elapse before cleaning is necessary.

It should not, of course, be necessary to

de-scale the second stage boiler on the

A4000D or when the stills have been

operated from a de-ioniser.

Cleaning the still involves the use of

concentrated hydrochloric acid. As a

safety precaution protective clothing,

gloves, mask and goggles should be

worn during cleaning.

Note:
The cabinet and perspex screen should

be cleaned using a dilute detergent

solution only.

It is possible to descale the Aquatr on

Water Still without dismantling the

glassware by following these instructions

in conjunction with Control of

Substances Hazardous to Health

regulations (COSHH) 1988.

Method

1 Push the CLEAN switch to the

CLEAN position – the switch

illuminates.

2 Observe that the heating element/s

has switched off. Allow boiling to

stop and water to cool.

3 Push ON/OFF switch to OFF

position.

4 Remove perspex screen.

5 Open the Rotaflo stopcock on the

level control WL48, allow the boiler

to empty to

3

/

4

’s of its full capacity,

and close stopcock.

6 Carefully add about 100ml. of

concentrated hydrochloric acid into

the glass funnel of the WL48.

7 Push the ON/OFF switch to the ON

position, the boiler will refill to its

operating level.

8

Push the ON/OFF switch to the OFF

position

.

9

Allow the chemical reaction within

the boiler to continue until all the

deposits have been removed.

Note – if scale has built up above

the operating level, the boiler may

be completely filled by either lifting

the drain pipe or restricting the

flow of water to drain. This

operation must be carefully

controlled so that the boiler does

not over-fill and cause water to be

forced over the funnel top.

10 Open Rotaflo stopcock and allow

contents of the boiler to drain

completely.

Note:

If the acid added to the boiler has

not been completely neutralised,

the liquid flowing to drain may be

strongly acid. Necessary safety

precautions should be observed

around the drain position, and any

effluent control procedures

followed.

11 Close Rotaflo stopcock.

12 Push ON/OFF switch to ON position

and allow boiler to fill with water.

Again the boiler may be completely

filled by following procedure

detailed in step 8.

13 Push ON/OFF switch to OFF

position.

14 Drain the boiler by opening the

Rotaflo stopcock..

15 Repeat steps 11-14 two or three

times until the boiler has been

thoroughly flushed through.

16 The still is re-started by finishing

the cleaning cycle at step 12, and

then pushing the CLEAN switch to

the OFF position. (Switch is not

illuminated).

Before collecting the distillate the

still should be allowed to run for

about 10 minutes with the distillate

running to drain, this will ensure

that any residual acid is removed.

Note:

The amount of acid required will

depend upon the degree of scaling.

If after performing the cleaning

cycle deposits are left in the boiler,

it may be necessary to repeat the

cleaning cycle using a fresh

quantity of acid.

The benefits gained from regular

cleaning cannot be over

emphasised.

Cleaning
A4000, A8000,
A4000D