17 Pass the electric cables of the

heaters up into the electrical

control compartment via the

available orifice. Push the spade

terminals onto the correct pins of

the connecting strip (refer to fig. 6).

18 Refit the cabinet end panel with all

six screws and washers and the

cabinet roof with all six

countersunk screws. Replace the

perspex viewing screen.

You have now assembled your

Aquatron Water Still. Move the

assembled A8000 to its working

location – using the mounting

bracket, item 7 if the unit is to be

wall mounted. Ensure there is a

space of at least 50cm to the right

of the unit to allow access to the

heating elements.

19 Connect the cold water supply to

connection 18 fig. 4. Note the

tubing selected should have a safe

working pressure of at least equal

to the pressure of the water supply

and should be adequately secured

by hose clips.

DO NOT TURN ON WATER
SUPPLY.

20 Electrical Installation

THIS EQUIPMENT MUST BE
EARTHED!

The electrical installation should

only be carried out by a qualified

electrician.

The A8000 is classed as

permanently connected equipment.

The equipment is supplied with

1.5m of flexible, triple core, circular

cable to the following

specification:- 4.0mm

2

, to BS 6500

or equivalent and <HAR> or BASEC

approved.

Connection to the mains electrical

supply, should be via a double pole

isolation switch with a continuous

current carrying capacity of 30A at

250v and overcurrent protection

should be provided by a double

pole approved fuse rated at 30A.

These devices should be sited near

to the equipment and clearly

marked – “Disconnect device for

Aquatron Water Still“

The colour code for the mains

cable is:–

Brown

– Live

Blue

– Neutral

Yellow/Green – Earth

Fig 6 Terminations for A8000

Mains Cable Replacement
If the mains cable requires

replacement, only specially

prepared spare mains lead obtained

from Bibby Sterilin should be used.

DO NOT SWITCH ON
ELECTRICITY SUPPLY.

21 Connect the reservoir level control

WLS as detailed on page 14 of the

manual.

22 Refer to page 9 for operating

instructions.

Blue - neutral
Brown - live
Earth - yellow/green

PCB

Thermostat

Relay

Relay

N

E

L

Heater

lead

(front)

Heater

lead

(back)