Dynalab A4000D Manuel d'utilisation

Page 14

Advertising
background image

[ 12 ]

Blue - neutral

Brown - live

Earth - yellow/green

Fig 9 Terminations for A4000D

PCB

Thermostat

(Rear)

Thermostat

(Front)

Relay

Relay

N

E

L

Heater

lead

(front)

Heater

lead

(back)

16 Refit the metal thermostat probe

into glass tube of the boiler.

17 Take the one metre length of

16mm tubing and insert through

the hole, item 22 in the cabinet

side. The tubing should extend

approximately 250mm into the

cabinet, roughly level with the

outlet of the condenser.

18 Identify the front boiler level

control WL48, item 27, fig.7, and

connect the outlet to the tubing.

Secure with a tie strap. Ensure that

the Rotaflo Stopcock of the WL48

is closed.

19 Couple the WL48 to the boiler

connection arm with the plastic

coupling and rubber sleeve, item

28. Ensure there are no stresses

from the 16mm tubing on the

connection.

20 Looking inside the cabinet, identify

the vinyl tubing assembly fitted

with 5 numbered screwthread

connectors. Referring to illustration

fig. 8 make the following

connections:–

Connector no. 1 to lower water

inlet of rear condenser.

Connector no. 2 to lower water

inlet of front condenser.

Connector no. 3 to upper water

outlet of rear condenser.

Connector no. 4 to upper water

outlet of front condenser.

Connector no. 5 to front boiler level

control WL48.

21 The tubing assembly is also fitted

with a 1 metre length of 9mm

diameter vinyl tubing. Feed the free

end of this tubing through the

opening, item 21, in the side of the

cabinet.

22 Unscrew the black plastic screwcap

from the WL48 and assemble the

glass funnel WF48 – item 4.

23 Unscrew the black plastic screwcap

and rubber ring from the rear boiler.

Slide both onto heating element

W48H – item 29.

24 Assemble the heater into the boiler

ensuring that the far end is

supported in the glass nodule of the

boiler.

Note: It is not necessary to over
tighten the screwcap and risk
breaking the glass boiler.

25 Pass the electric cable of the heater

up into the electrical control

compartment via the orifice

available. Push the spade terminals
onto the CORRECT PINS of the
connecting strip (refer to fig. 9).

26 Repeat the operation to fit the

second heating element for the
front boiler.

27 Refit the cabinet end panel with all

six screws and washers. Refit
cabinet roof with the six
countersunk screws.

28 Refit the perspex viewing panel.

29 Move assembled A4000D to

desired working location. ensure
there is a space of at least 50cm to
the right of the unit to allow access
to the heating elements.

30 Connect the cold water supply to

connection 18 fig. 7.

Note:

The tubing selected should

have a safe working pressure at
least equal to the pressure of the
water supply. The tubing should be
adequately secured by hose clips.

DO NOT SWITCH ON.

31 Electrical Installation

THIS EQUIPMENT MUST BE
EARTHED!

The electrical installation should
only be carried out by a qualified
electrician.

The A4000D is classed as
permanently connected equipment.

The equipment is supplied with
1.5m of flexible, triple core, circular
cable to the following
specification:- 4.0mm

2

, to BS 6500

or equivalent and <HAR> or BASEC
approved.

Connection to the mains electrical
supply, should be via a double pole
isolation switch with a continuous
current carrying capacity of 30A at
250v and overcurrent protection
should be provided by a double
pole approved fuse rated at 30A.

These devices should be sited near
to the equipment and clearly
marked – “Disconnect device for
Aquatron Water Still“

The colour code for the mains
cable is:–

Brown

– Live

Blue

– Neutral

Yellow/Green – Earth

Mains Cable Replacement
If the mains cable requires
replacement, only specially prepared
spare mains lead obtained from Bibby
Sterilin should be used.

DO NOT SWITCH ON
ELECTRICITY SUPPLY.

32 Connect the reservoir level control

WLS as detailed on page 14 of the
manual.

33 Refer to page 13 for operating

instructions.

Fig 10
W4L2A
Level Control

To boiler level

socket

Guide

tube

Float switch

Distillate feed

from front still

To drain

15 - 20mm