41

IM 12D08N05-01E

9

QUALITY

INSPECTION

2/3

QIS 12D08N05-01E

This test is done on the “HIF” display of “Factory Mode”.

a. Touch the [Setup] icon.

b. Touch the [Commissioning].

c. Touch the [Advanced setup].

d. Touch the [Factory adjustment].

e. Enter a password.

f. Select the [Factory Mode] in “Key.”

g. Select the [HIF] in “Execute.”

When the resistance of the decade resistance box 2 to the corresponding value

“RANGE1” to “RANGE5” in Table 1 is set, check the data display and the value must be

within the range shown in Table 1.

Table 1

RANGE

Resistance (Ω) of

Decade Resistance Box 2

Data Display (Ω)

RANGE1

20

20 ±0.1

RANGE2

200

200 ±1

RANGE3

2k

2k ±0.01k

RANGE4

20k

20k ±0.1k

RANGE5

200k

200k ±1k

After the above test is completed, touch the [Exit] to return to the “HIF” display.

3.4 Temperature Indication Check

Following Section 3.3, select the [PT1000] in “Temperature” of the “HIF” display.

In this state, change the resistance of the decade resistance box 1 and check the data

display. The value on the data display must be within the range shown in Table 2.

Table 2

Temperature (°C)

Resistance (Ω) of

Decade Resistance Box 1

Data Display (°C)

–10

960.9

–10 ±0.3

25

1097.3

25 ±0.3

240

1904.6

240 ±0.3

After the above test is completed, touch the [Exit] to return to the “HIF” display.

3.5 Current Output Test

Following Section 3.4, select the [Check] in “mA outputs” of the “HIF” display. “Set value

4.000 mA” appears at the bottom of the display. Select “Next value” in the “Command”

and touch “Enter,” the value on the data display increases in steps of 4 mA. Check the

current outputs 1 and 2 corresponding to the data display, the current output must be

within the range shown in Table 3.

Table 3

Data Display

Current Output (mA DC)

4

4 ±0.02

12

12 ±0.02

20

20 ±0.02