Yokogawa EXAxt SC450 4-Wire Analyzer for Conductivity/Resistivity Manuel d'utilisation

Page 32

Advertising
background image

25

IM 12D08N05-01F-E

5

STRUCTURE

DES

MENUS

Menu

Paramètre

Valeurs

Etendue

par défaut

min.

max.

mA1 (control)

Expire time

0.0 sec.

0 sec.

1800 sec.

mA1 (output)

Damping time

0.0 sec.

0 sec.

3600 sec.

mA1 (simulate)

Simulation perc.

50%

0%

100%

PID-control mA1

Setpoint

250.0 µS/cm

- inf

+ inf

PID-control mA2

Setpoint

25ºC/ºF

- inf

+ inf

PID-control mA1

Range

50.00 µS/cm

- inf

+ inf

PID-control mA2

Range

10ºC/ºF

- inf

+ inf

PID-control mA1

Manual Reset

0%

0%

100%

PID-control

I-time

3600 sec.

1 sec.

3600 sec.

PID-control

D-time

0 sec.

0 sec.

60 sec.

Linear mA1

0% Value

0 S/cm

- inf

+ inf

Linear mA2

0ºC/ºF

- inf

+ inf

Linear mA1

100% value

500.0 μS/cm

- inf

+ inf

Linear mA2

100ºC/ºF

- inf

+ inf

Table

Table mA1

voir annexe

- inf

+ inf

mA2 semblable à mA1