6 print menu, 7 initial scale setup (cont.) 2.7.6 print menu, Procedure – Ohaus CD-11 Indicator Manual en Manuel d'utilisation

Page 16: Cd-11 indicator

Advertising
background image

16

2.7 Initial Scale Setup (Cont.)
2.7.6 Print Menu

The Print menu provides data communication settings which can be entered. It contains 6 submenus: Reset, Baud
rate, Parity Bit, Data Length, Stop Bits and End/Store.

Procedure

To select any of the items in the Print menu, proceed as
follows: NOTE: If you have entered from the preceeding
menu, disregard the first step.

• With the Indicator ON, press and hold the Zero/Menu

button until MENU is displayed. When you release
the Zero/Menu button, CAL is displayed, then press
Print/Units button, until Print is displayed.

• Press Zero/Menu button, rESEtn is displayed. This

allows resetting the Print menu to factory defaults.
rESETn = no does not reset settings. rESETy= yes
will reset the entire Print menu as follows:
Baud rate =2400, parity =none, data length=7,
stop bit=2.

• Press Print/Units button, and select N or Y.

• Press Zero/Menu button, bd2400 displayed.

• Press Print/Units button, and select desired baud

rate. Baud rate selections are: 1200, 2400, 4800 and
9600. 2400 is the default setting.

• Press Zero/Menu button, PAr NO is displayed. This

is the parity bit.

• Press Print/Units button, and select desired parity of

NO=none, Odd=odd, E=even. Default setting is none.

• Press Zero/Menu button, dAtA 7 is displayed. This

is the data length.

• Press Print/Units button, and select desired data

length of 7 or 8. Default setting is 7.

• Press Zero/Menu button, StOP 2 is displayed. This

is the stop bit.

• Press Print/Units button, and select desired stop bit of

1 or 2. Default setting is 2.

• Press Zero/Menu button to end this block, END is

displayed.

• Press Zero/Menu button, LOCSW is displayed which

is the next menu or press Print Units button to go back
to the Print menu.

(If initial setup, go to the next paragraph. To exit from
the Setup, press Prints/Units button to skip to PRINT
then to LOCKSW, then QUIT. Press Zero/Menu
button to go back to the weighing mode).