2.7 Initial Setup (Cont.)
2.7.4 Setup Menu

The CD-11 Indicator Setup Menu

must be entered the first time the Indicator is used to set the scale base param-

eters to match the Indicator. Do not attempt to calibrate the Indicator before setting up the Setup Menu. All other
menus should be entered and set up the first time the Indicator is used.

Procedure

• With the Indicator ON, press and hold the Zero/Menu

button until MENU is displayed. When you release
Zero/Menu button, CAL is displayed when the CAL
jumper on the PC board is in place. When the CAL
jumper is removed, the Indicator will not permit cali-
bration. This jumper should be in place initially.

• Press Print/Units button, SETuP is displayed.

• Press Zero/Menu button, LFTOFF is displayed. legal

for trade selections are:

'ON' - LFT is ON
'OFF' - LFT is OFF.

• Press Print/Units button and select either ON or

OFF.

• Press Zero/Menu button, 0 2 is displayed. This is

the Zero 2%, 18% or 100% setting. 2% - zero opera-
tion range is - 2% to + 2%. 18% - zero operating
range is

-2% to +18%, 100% - zero operation range

is -2% to +100%. NOTE: If LFT is ON, only 2% and
18% are available.

• Press Print/Units button, and select either 2%, 18% or
100%.

• Press Zero/Menu button, CAL Un kg is displayed.

This is the calibration unit setting. Selections are:

'lb' - calibration unit is lb
'kg' - calibration unit is kg.

• Press Print/Units button, and select either kg or lb.

• Press Zero/Menu, F xx is displayed. This is full

scale capacity selections. xx= value last set.
Selections are:

5, 10, 20, 25, 30, 40, 50, 60, 75, 100, 120, 200, 250,

300, 400, 500, 600, 750, 1000, 1500, 2000, 2500,
3000, 5000, 7500, 10000, 20000 (lb or kg).

• Press Print/Units button until desired capacity value

is reached.

• Press Zero/Menu button, Gd0.01 is displayed. This

is the graduation size. Refer to paragraph 2.7.3 Load
Cell Capacity Information table. For available selec-
tions, press Print/Units button until desired gradua-
tion value is reached.

• Press Zero/Menu button, CP 30 kg is displayed.

This is the full scale calibration point setting. The
range is from 20% to 100% Full scale capacity. Press
Print/Units button until desired calibration value is
reached.

• Press Zero/Menu button to end this block, END is

displayed.

• Press Zero/Menu button, rEAD is displayed which is

the next menu or press Print/Units button to return to
Setup menu.

The Indicator is now matched up with the scale base and
the Indicator parameters may now be set and calibrated.