Ohaus CD-11 Indicator Manual en Manuel d'utilisation

Page 15

Advertising
background image

15

AUTO POWER OFF

• Press Zero/Menu button, AOtOFF is displayed. This

is the Auto Off Timer. When set ON, the Indicator will
shut off automatically after 5 minutes has elapsed on
the condition that no button is pressed and the scale
base is stable during that period.

• Press Print/Units button,and select ON or OFF.

OFF is the default setting.

RETAIN ZERO DATA

• Press Zero/Menu button, Un rZdOFFis displayed.

This is Retain Zero Data which can be turned on or
off. When set On, the Indicator stores the current
zero point and restores it on the power-up.

• Press Print/Units button, and select ON or OFF.

OFF is the default setting.

• Press Zero/Menu button to end this block, END is

displayed.

• Press Zero/Menu button, Print is displayed which is

the next menu or press Print/Units button to go back
to Readout menu.

(If initial setup, go to the next paragraph. To exit from
the Setup, press Prints/Units button to skip to PRINT
then to LOCKSW, then QUIT. Press Zero/Menu
button to go back to the weighing mode).

2.7 Initial Setup (Cont.)
2.7.5 Readout Menu (Cont.)

Procedure

CD-11 Indicator