13

Sim

ulation Softwar

e

Sim

ulator Setup

Maintenance

Spar

e Par

ts

Specifications

Intr

oduction

Tr

oubleshooting

Maintenance

Opening the Torso

Open the Simulator torso to perform procedures such as:

– Attaching or replacing limbs

– Replacing the Simulator Battery

– Replacing the Chest Rise Bladder and Lung Bladder

– Replacing the Torso Skin

– For general inspection

Warning: Open the torso from left to right, gently unhook the

skin. Do not pull hard to lift the Torso Skin, this could damage
the chest cables.

1. Unhook Torso Skin holders on left side of the torso and each

shoulder.

Replacing the Simulator Battery

1. Open the Simulator torso.

2. Fold the Torso Skin over to the right side.

3. Disconnect the Defibrillation connector cable.

4. The Defibrillation connector cable must be disconnected at the

right side of the Simulator below the chest form.

Note: DO NOT disconnect the tubes and cables connecting
the stomach foam to the Simulator.

5. Replace the Torso Skin by performing steps 1- 4 in reverse.

2. Slide the battery out of the battery strap.

3. Pull the battery power connection up from inside the

Simulator. Unhook the battery power connection.

4. Lift the battery out of the Simulator.

5. Replace the charged battery or install a new Laerdal

approved SimJunior battery.

6. Connect the battery to the Simulator.

7. Replace the Torso Skin.