Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP Manuel d'utilisation

Page 34

Advertising
background image

28

IM 12B6B5-F-E

Menu

Paramètre

Défaut

Etendue

min.

max.

Régul.PID S1

Consigne

7pH

-inf

+inf

Régul. PID S1

Etendue

1.00 pH

0.001 pH

+inf

Régul. P(ID) S1

Manuel

0%

0%

100%

Régul. PI(D) S1

Temps I

3600 sec.

1 sec.

3600 sec.

Régul. P(I)D S1

Temps D

0 sec.

0 sec.

60 sec.

Cycle

Temps DC

10 sec.

1 sec.

1800 sec.

Pulse freq.

Max. pulse freq.

70 p/min.

1 p/min.

70 p/min.

mA1 (simuler)

Temps d’expiration

0.0 sec.

0 sec.

1800 sec.

Alarme S1

Consigne

13 pH (high)

-inf

+inf

Alarme S2

Consigne

1 pH (low)

-inf

+inf

Alarme

Hystérésis

0.10 pH

0 pH

+inf

Alarme

Temps de retard

0.2 sec.

0 sec.

1800 sec.

Hold

Valeur fixe mA1

12 mA

3.6 mA

21 mA

Hold

Valeur fixe mA2

12 mA

3.6 mA

21 mA

(réguler)

Paramètre process

Réglage Régulation P
Type de Cycle
régulation
Tps d’expiration 0.0 S
(0 S: désactivé)

Régulation P

Régulation PI
Régulation PID

Fréquence impuls

Cycle

Consigne
Bande

bande proportionnelle

Temps I

Direction Inverse

Temps D

Maintien
Wash
Défaut
Simuler
Non actif

Alarme

Régulation

Paramètre process

(alarme)

Réglage alarme

bas

Inverse

haut

Point consigne
Direction
Hystérésis
Temps retard
Tps expi.
0 s= désactivé

Réglage de la sortie

Configuration du maintien

Sortie

Sortie

S2, S3, S4 identique à S1

Maintien

Dernière ou fixe
Valeur fixe mA1
Valeur fixée mA2

Les contacts de sortie sont
réglés inactifs durant maintien
Maintien durant

Activé

calibration

Actif

Non actif

Dernière

Fixe