3.1 General rules

794 Basic Titrino

81

.Req.Buf2:

Instrument in the CAL mode, requesting pH of buffer 2.

.Meas.Buf2:

Instrument in the CAL mode, measures buffer 2.

etc.

$G .Assembly.Bur .Fill:

Buret in filling process

.ModeDis:

Buret in DIS mode

In TIP, its global status as well as the step number (X) is available.

$G .TIP.X

.Inac:

Instrument at the beginning or at the end of a TIP.

.Req .Id1:

Instrument in the TIP mode, requesting Id1 after start.

.Id2:

Instrument in the TIP mode, requesting Id2 after start.

.Id3:

Instrument in the TIP mode, requesting Id3 after start.

.Smpl:

Instrument in the TIP mode, requesting sample size after start.

.Unit:

Instrument in the TIP mode, requesting unit of sample size after start.

.Pause:

Instrument in the TIP mode, in pause.

.Info:

Instrument in the TIP mode, in info.

.Mode...:

Instrument in the TIP mode, working off a submethod. The detailed

status messages of the submethod appear, see above.

Status conditions of the global $H:

The status message of the action which has been held appears.

If the process is held because a monitored limit has been violated, its status

message is $H.Mode.XXX.Titr.

Status conditions of the global $C:

The status conditions of the global $C are identical with the ones of the global

status $G. They appear when the process has been restarted actively from the

status "Hold" ($C, key <meas/hold> or automatically after elimination of an

error).

Status conditions of the global $R:

$R .Mode.XXXX.QuickMeas:

Quick manual measurement from the initial status in mode XXXX.

$R .Mode.DET .Inac:

Instrument in the DET mode, inactive.

$R .Mode.MET .Inac:

Instrument in the MET mode, inactive.

$R .Mode.SET .Inac:

Instrument in the SET mode, inactive.

.Cond.Ok:

Instrument in the SET mode, conditioning, endpoint reached.

.Cond.Prog:

Instrument in the SET mode, conditioning, endpoint not reached.

$R .Mode.MEAS .Inac:

Instrument in the MEAS mode, inactive.

$R .Mode.CAL .Inac:

Instrument in the CAL mode, inactive.

$R .Assembly.Bur.ModeDis:

Buret in the DIS mode, inactive.

$R .TIP.Inac:

Instrument in TIP, inactive.

Status conditions of the global $S:

$S .Mode.XXXX.QuickMeas:

Quick manual measurement from the initial status in mode XXXX.

The instrument gives the status from which it has been stopped. The detailed

status information is therefore identical to for the global status $G.

Violation of monitored limits with action "end" give the status message

$S.Mode.XXX.Inac;EYYY.

Advertising