Metrohm 794 Basic Titrino Manuel d'utilisation

Page 113

Advertising
background image

3.2 Remote control commands

794 Basic Titrino

109

3.2.2.36. Mode.Parameter.Calibration.CalTemp -20.0...25.0...120.0

Parameters for CAL: Calibration temperature in °C. If a Pt 100 or Pt1000 is

connected, the temperature is measured.


3.2.2.37. Mode.Parameter.Calibration.Buffer.1.Value 0...7.00...±20.00

Mode.Parameter.Calibration.Buffer.2.Value 0...4.00...±20.00, OFF

etc. up to 9 buffers

Parameters for CAL: pH of buffers. The first buffer which is set to "OFF" de-

termines the number of buffers in the calibration.

3.2.2.38. Mode.Parameter.Calibration.SignalDrift 0.5...2...999, OFF

Mode.Parameter.Calibration.EquTime 0...110...9999, OFF

Parameters for CAL: Criteria for measured value acquisition. Measured value

drift in mV/min, equilibration time in s. OFF means that the corresponding cri-

terion is switched off. If both criterions are on OFF, the measured value is ac-

quired immediately.

If the equilibration time has never been edited, it is automatically calculated by

the instrument to match the drift, see page 31. After it has been edited once, it

remains in force with the set value.

3.2.2.39. Mode.Parameter.Calibration.ElectrodeId

up to 8 ASCII char

Parameters for CAL: Electrode identification. It is classified under calibration

data, see 3.2.2.68.

3.2.2.40. Mode.Parameter.Calibration.SmplChanger ON, OFF

Parameters for CAL: Calibration at Titrino.

With "ON", there are no hold points in the calibration sequence for entries, the

first buffer is measured directly.

3.2.2.41. Mode.Parameter.Calibration.ActPulse first,

all, OFF

Parameters for CAL: Output of a pulse on the I/O line "Activate", see page

163.

3.2.2.42. Mode.Parameter.Sequence.X.Select

method, pause, L4 output,

L6 output, info, OFF

Parameters for TIP: Selection of an element for step X (X = 1...30). For the

parameters of the elements see 3.2.2.43.

3.2.2.43. Mode.Parameter.Sequence.X.Method Method

name

Mode.Parameter.Sequence.X.Pause

0...999 999, INF

Mode.Parameter.Sequence.X.L4Output

active, inactive,

pulse, OFF

Mode.Parameter.Sequence.X.L6Output

as for L4

Mode.Parameter.Sequence.X.Info

up to 16 ASCII characters

Advertising