3.2 Remote control commands

794 Basic Titrino

123

.P

PowerOn: Simulation of power on (3.2.2.99). Not from mains.

Messages from node .T, Titrator:

.T.R

Ready: Status 'Ready' has been reached.

.T.G

Go: Instrument has been started.

.T.GC

GoCommand: Instrument (or submethod in TIP) has received a go

command.

.T.S

Stop: Status 'Stop' has been reached.

.T.B

Begin of sequence (or submethod).

.T.F

Final: End of determination (or submethod), the final steps will be

carried out.

.T.E

Error. Message together with error number, see page 82ff.

.T.H

Hold: Status 'Hold' has been reached.

.T.C

Continue: Continue after hold.

.T.O

Conditioning OK: EP reached (in SET with conditioning).

.T.N

Conditioning Not OK: EP not reached (in SET with conditioning).

.T.Re

Request: In the inquiry of an identification or the sample size after

start of titration.

.T.Si

SiloEmpty: Silo empty, i.e. the last line has been removed from

the silo memory.

.T.M

MeasList: Entry in the measuring point list (with DET, MET).

.T.EP

EPList: Entry into EP list (with DET, MET, SET)

.T.RC

Results have been recalculated.

Messages for changings in the I/O lines. If the changings are made simulta-

neously, there is 1 message. Pulses receive 2 messages: one message each

for line active and inactive.

.I

Input: Change of an input line.

.O

Output: Change of an output line (except 7, pin 2, for recorder

pulses).

If a change occurs that requires a message, the Titrino sends space (ASCII

32) and ! as an introducer. This is followed by the name of the device (see

3.2.2.55). Special ASCII characters in the device name are ignored. If no de-

vice name has been entered, only ! is sent. Finally the Titrino sends the infor-

mation which node has triggered the message.

Example: !John".T.Si": The message was triggered from instrument "John",

node .T.Si

3.2.2.98. Setup.Graphics.Grid ON, OFF

Setup.Graphics.Frame ON, OFF

Setup.Graphics.Scale Full, Auto

Setup.Graphics.Recorder.Right 0.2...0.5...1.00

Setup.Graphics.Recorder.Feed 0.01...0.05...1.00

Change in the appearance and the format of the curve for the output. The set-

tings are valid for both Titrino COM ports.

.Grid:

On/off switching of grid over curve.

.Frame:

On/off switching of frame surrounding the curve. If grid and frame

are switched off, the curve is printed faster as the print head does

not have to move to the end of the paper.

.Scale:

Type of scaling of the measured value axis: Full means that the

scale runs from the smallest up to the greatest measured point.

With auto, the smallest measured value is taken and the next

smaller tick defines the beginning of the scale; the next greater

tick to the greatest measured value is the end of the scale.