Metrohm 794 Basic Titrino Manuel d'utilisation

Page 112

Advertising
background image

3.2. Remote control commands

794 Basic Titrino

108

3.2.2.30. Mode.Parameter.TitrPara.Direction +,

-, auto

Parameters for SET: Titration direction.

"auto" means the titration direction is determined automatically by the instru-

ment. If 2 EP's have been set in a SET titration, the titration direction is given

by the two EP's. The entry of the titration direction is then invalid.

3.2.2.31. Mode.Parameter.Presel.Cond ON, OFF

Mode.Parameter.Presel.DriftDisp ON, OFF

Parameters for SET:

.Cond: Conditioning ON/OFF

.DriftDisp: Drift display during conditioning ON/OFF.

3.2.2.32. Mode.Parameter.Measuring.SignalDrift

pH, U, Ipol, T:

0.5...999, OFF

Upol: 0.05...99.9, OFF

Mode.Parameter.Measuring.UnitSigDrift read

only

Mode.Parameter.Measuring.EquTime 0...9999, OFF

Parameters for MEAS: Criteria for the measured value acquisition. Measured

value drift in mV/min (with pH, U, Ipol), µA/min (with Upol), resp. °C/min

(with T). Equilibration time in s. OFF means that the corresponding criterion is

switched off. If both criteria are OFF, the measurement continues indefinitely.

If the equilibration time has never been edited, it is automatically calculated by

the instrument to match the drift, see page 45. After it has been edited once, it

remains in force with the set value.

3.2.2.33. Mode.Parameter.Measuring.MeasInput

1, 2, diff.

Mode.Parameter.Measuring.Ipol ±127...1...+127

Mode.Parameter.Measuring.Upol ±1270...400...+1270

Mode.Parameter.Measuring.PolElectrTest ON, OFF

Parameters for MEAS:

Selection of the measuring input; valid with measured quantities pH and U.

"diff." means differential amplifier, see page 155.

With Ipol, the inquiries for the polarization current in µA (Ipol) and

.PolElectrTest are valid.

With Upol the inquiry for the polarization voltage in mV (Upol) is valid. Entry in

steps on 10 mV.

Besides .PolElectrTest is valid.

If the test for polarized electrodes is switched on, it is performed on change-

over from the inactive state to the measurement.

3.2.2.34. Mode.Parameter.Measuring.Temp -170.0...25.0...500.0

Parameters for MEAS: Measurement temperature in °C. If a Pt100 or Pt1000

is connected, the temperature is measured.

The temperature is used for the temperature correction in pH measurements.

3.2.2.35. Mode.Parameter.Calibration.MeasInput

1, 2, diff.

Parameters for CAL: Selection of the measuring input. "diff." means differen-

tial amplifier, see page 155.