3.2 Remote control commands

795 KFT Titrino

89

.Feed:

Length of the time axis. Depending on the printer, the measure in

cm may not always be correct.

Curve

length

0.01 100

cm

0.1 10

cm

0.5 2

cm

1 1

cm

3.2.2.94. Setup.PowerOn $G

Simulation of 'power on'. The device has the same status as after power on:

The cylinder is filled, error messages deleted and the current sample number

set to 0. The method last used is ready for operation.

3.2.2.95. Setup.Initialise $G

Setup.Initialise.Select ActMeth, Silo, Config,

Assembly, Setup, All

Setting of default values for the following areas:

ActMeth: Current method. Parameters, calculations, and assignments for

the data output, operands C01...C19.

Silo:

The silo memory is deleted. Same function as delete entire silo.

Config:

All values under &Config.

Assembly: All values under &Assembly.

Setup:

All values under &Setup.

All:

Values of the entire tree (except silo and method memory).

The action must be triggered with &Setup.Initalise $G.

3.2.2.96. Setup.RamInit $G

Initializes instrument, see page 110. All parameters are set to their default

value and error messages are cleared. The user and silo memories will be de-

leted. The user memory contains the default user methods from Metrohm.

3.2.2.97. Setup.InstrNo $G

Setup.InstrNo.Value serial

number, 8 ASCII characters

Instrument identification for report output.

Set the value with &Setup.InstrNo $G .

3.2.2.98. Diagnose.Report $G

Output of the report containing the adjustment parameters. The Titrino has to

be in its inactive basic state.