3.2 Remote control commands

795 KFT Titrino

68

The limit control for results can also be activated. If a result is out of limit, a

message appears in the result report, E196 is sent, and output line L13 can

be set.

3.2.2.6. Mode.Def.SiloCalc.Assign.C24

RSX, EPX, CXX

Mode.Def.SiloCalc.Assign.C25

RSX, EPX, CXX

Mode.Def.SiloCalc.MatchId

id1, id1&2, all, OFF

.Assign.C2X: Assignment to store results in the silo as C2X.

.MatchId:

Indication which sample identification(s) have to match so that

the results can be combined.

3.2.2.7. Mode.Def.ComVar.C30

RSX, MNX, EPX, CXX

Mode.Def.ComVar.C31

etc., up to .C39

Assignment of common variables.

The values of the common variables are to be found in &Config.ComVar. They

can be viewed and entered there, see 3.2.2.50.

3.2.2.8. Mode.Def.Report.Assign1

Mode.Def.Report.Assign2

param, full, short, mplist, curve, scalc full, scalc srt, calc, ff

Definition of the report sequence, which is outputted automatically at the end

of the determination. Entries of more than one block have to be separated with

";".

.Assign1: Output to COM1 of the Titrino. Identical for COM2.

3.2.2.9. Mode.Def.Mean.1.Assign RS1, RSX, EPX, CXX

Mode.Def.Mean.2.Assign

etc., up to .9

Assignment of the statistics calculations. Valid assignments are a requirement

for statistics calculations. In addition, the statistics calculation must be swit-

ched on, see 3.2.2.24. Rules for statistics calculations see page

29.

3.2.2.10. Mode.CFmla

Mode.CFmla.1.Value

0...±999999

Mode.CFmla.2.Value

etc., up to .19

Calculation constants specific to a method. Stored in the method memory of

the Titrino. Operands specific to the sample (3.2.2.52 and 3.2.2.63) and val-

ues of common variables (3.2.2.50) on the other hand are not stored with the

methods.