3.2 Remote control commands

795 KFT Titrino

72

3.2.2.26. Mode.Parameter.Presel.IReq

id1, id1&2, all, OFF

Mode.Parameter.Presel.SReq

value, unit, all, OFF

Automatic inquiry after the start of the determination. From such an inquiry,

the determination continues if the requested entry/entries is/are made, e.g.

&SmplData.OFFSilo.Id1 (see 3.2.2.52) or with &M $G, see 3.2.2.1.

$H is not possible in requests.

3.2.2.27. Mode.Parameter.Presel.LimSmplSize.Status ON,

OFF

Mode.Parameter.Presel.LimSmplSize.LoLim

0.0...999 999

Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999

Limit control for the sample size.

3.2.2.28. Mode.Parameter.Presel.Oven COM1,

COM2,

no

If an oven is connected, its result will be incorporated into the result report of

the Titrino. If there is no oven connected via RS232, the setting of this pa-

rameter has to be "no".

3.2.2.29. Mode.Parameter.Presel.ActPuls

first, all, cond., OFF

Output of a pulse on the I/O line "Activate", see page 126.

3.2.2.30. UserMeth.FreeMem read

only

Memory space, available for user methods or silo lines. $Q sends the number

of free bytes, e.g. "4928".

3.2.2.31. UserMeth.Recall $G

UserMeth.Recall.Name

up to 8 ASCII characters

UserMeth.Store $G

UserMeth.Store.Name

up to 8 ASCII characters

UserMeth.Delete $G

UserMeth.Delete.Name

up to 8 ASCII characters

UserMeth.DelAll $G

Management of the internal method memory: Load, store and delete methods.

An action is performed if "$G" is sent to the corresponding node just after en-

tering the name.

Do not use blank characters before and after method name!

.DelAll: Deletes all methods in the user memory.